Oven Incubators

General Purpose Incubators



50 litre Incubator with Digital Control INC/50/DIG


75 litre Incubator MINI/75


The GENLAB General Purpose Incubators offer a range of highly efficient, reliable, cost effective units to suit most biological analysis, research and general laboratory applications.

Specifications

- 10 sizes 6 to 200 litres

-Temperature range:
ambient +5 to 100ºC without fan
ambient +5 to 70ºC with double door
ambient +5 to 60ºC with window
ambient +8 if fan fitted

- Fluctuation +/- 0.25ºC @ 37ºC

- Easy clean powder coated body

- Aluminium coated mild steel chamber

-Direct reading thermostat

-Safety overheat thermostat to DIN 12-880, Class 3.1

-Full two year warranty

C.E. compliant
Options

- Stainless steel chamber

- Fan circulation (30 litre +)

-Alternative door with integral viewing window

- Double door (solid outer, clear inner)

-Microprocessor digital controller with dual display of set point and actual temperature

-Other accessories available on request

Design
The exterior is constructed from sheet steel finished in an easy clean powder coated paint. The interior chamber is made from mild steel coated with aluminium (CLAD) with a stainless steel chamber available as an option. Fitted with fixed shelf runners and removable chrome plated wire grid shelves. The top vent is fitted with a clip to hold a glass immersion thermometer. The vertical style units are fitted with the controls below the door, the horizontal style units with the controls fitted on the side. The incubators have a solid steel door as standard or as an alternative a door with an integral viewing window (VIS) can be fitted. As an option a double door, comprising steel outer with an inner glass door is also available.
Heating
Heated by Incoloy sheathed elements; positioned below the chamber floor for natural convection units and fitted around the fan on the back or side wall of the chamber for mechanical convection units. The 200 litre units have fan circulation as standard.
Controls
The control system comprises of a direct reading thermostat and overheat thermostat both with calibrated scales and tamper proof locks. They also include main switch with indicator and heat and overheat indicators.

Options include a PID Microprocessor controller with dual displays of set point and actual temperature, which are auto-tuned for each individual unit to optimise the heat up, overshoot and control of temperature.